SX-COA01

SIGNUX COA01

Conductivity in one piece, ready for the bus

Integrated conductivity transmitter for continuous monitoring of water and aqueous solutions. Electrode and electronics in a single IP68 body, RS485 Modbus RTU output, automatic temperature compensation and salinity and TDS conversion. Two cell constants: k = 1 for 1-2000 µS/cm and k = 10 for 10-20000 µS/cm. 7-30 V DC.

This product adopts an integrated design, the structure is lighter andsimpler, and the use is more convenient. Waterproof grade IP68. It can be widely used in continuous monitoring of the conductivity value of aqueous solutions such as section water quality, aquaculture, sewage treatment, environmental protection, pharmaceuticals, food and tap water.


Measures
Conductivity · Temperature · Salinity · TDS
Output
Protection
IP68
Operating
-20 to +60 °C
SIGNUX COA01 integrated conductivity transmitter on a white background: slim black body with a cable gland on top and the electrode at the tip

What sets it apart

Maximum range of conductivity measurement

1~20000μs/cm

Tmperature measurement range

-20~60℃

Waterproof grace

IP68

Salinity and TDS included

Besides conductivity and temperature, the probe reports salinity and total dissolved solids in ppm, ready to use.

RS485 communication interface

communication address can be set, baud rate can be modified

The equipment adopts wide voltage power supply

DC 7~30V can be used

Configurations

Measurement rangePart number
k = 1 · 1-2000 µS/cmSX-COA01-1
k = 10 · 10-20000 µS/cmSX-COA01-10

Specifications

General

Measured parameters
Conductivity · Temperature · Salinity · TDS
Output
RS485Modbus RTU
Process connection
NPT 3/4"At both ends

Conductivity

Range
1-2000 µS/cmWith cell constant k = 1
10-20000 µS/cmWith k = 10
Resolution
0.1 µS/cmk = 1
1 µS/cmk = 10
Accuracy
±1 % FS

Temperature, salinity and TDS

Temperature range
-20 to +60 °C
Temperature resolution
0.1 °C
Temperature accuracy
±0.5 °C
Temperature compensation range
-20 to +60 °CAutomatic, referred to 25 °C; coefficient 0.02 by default, adjustable
Temperature compensation coefficient
0.02default

Other parameters

Salinity range
0-11476 ppmCalculated from conductivity
TDS range
0-13400 ppmCalculated from conductivity
Power supply
7-30 VDC
Power consumption
0.4 W
Configuration
Over RS485Address, baud rate, offsets, compensation coefficient, cell constant, calibration

Output and power

Output interface
RS485Modbus RTU, baud rate configurable from 1200 to 115200 baud
Default communication
4800 baud · address 18 data bits, no parity, 1 stop bit
Power supply
7-30 VDC
Power consumption
0.4 W
Configuration
Over RS485Address, baud rate, offsets, compensation coefficient, cell constant, calibration

Construction

Protection rating
IP68
Operating temperature
-20 to +60 °C
Maximum pressure
6 bar0.6 MPa
Dimensions
266 × ⌀30 mmBody 212 mm; electrode window 28 mm
Cable length
5 mOther lengths on request

Manual

1Introduction

The SX-COA01 is an integrated conductivity transmitter for aqueous solutions. Electrode, temperature sensor and electronics share one body, so the probe connects directly to an RS485 bus with Modbus RTU protocol and reports conductivity, temperature, salinity and total dissolved solids.

Temperature compensation is automatic: the conductivity read is referred to a reference temperature (25 °C by default) with an adjustable coefficient. Two cell constants are available, k = 1 for 1-2000 µS/cm and k = 10 for 10-20000 µS/cm.

Typical applications: surface water quality, aquaculture, wastewater treatment, environmental monitoring, pharmaceutical and food processes, drinking water.

  • Conductivity up to 20000 µS/cm; temperature -20 to +60 °C with 0.1 °C resolution
  • One-piece design, light and simple, easy to install
  • IP68 protection rating, pressure-rated to 0.6 MPa
  • Salinity and TDS conversion on board
  • RS485 interface with Modbus RTU protocol, easy to connect to a PLC, datalogger or computer
  • Configurable Modbus address and baud rate
  • Wide supply voltage range: 7-30 V DC

Specifications

Power supply7-30 V DC
Power consumption0.4 W
Communication interfaceRS485, Modbus RTU 4800 baud by default (1200, 2400, 4800, 9600, 19200, 38400, 57600 and 115200 selectable)
Conductivity rangek = 1: 1-2000 µS/cm, resolution 0.1 µS/cm k = 10: 10-20000 µS/cm, resolution 1 µS/cm
Conductivity accuracy±1 % FS
Temperature range-20 to +60 °C, resolution 0.1 °C
Temperature accuracy±0.5 °C
Temperature compensation-20 to +60 °C, referred to 25 °C by default; coefficient 0.02 by default
Salinity range0-11476 ppm
TDS range0-13400 ppm
Operating conditionsAmbient -20 to +60 °C
Protection ratingIP68
Pressure resistance0.6 MPa
Cable length5 m by default (other lengths on request)
SIGNUX COA01 integrated conductivity transmitter on a white background: slim black body with a cable gland on top and the electrode at the tip

2Wiring

The probe is supplied with a four-core cable with bare ends.

FunctionWire colourDescription
Power supplyBrownPower supply positive (7-30 V DC)
Power supplyBlackPower supply negative (GND)
CommunicationYellowRS485 A
CommunicationBlueRS485 B

3Dimensions and ordering information

3.1Dimensions

Dimensioned drawing of the COA01: 266 mm overall, 212 mm body, 30 mm diameter, 28 mm electrode window and NPT 3/4 thread at both ends
Overall length 266 mm, body 212 mm, diameter 30 mm, electrode window 28 mm, NPT 3/4" thread at both ends

3.2Ordering information

The order code states the cell constant. It cannot be changed afterwards: choose it from the conductivity to be measured.

Order codeCell constantRangeResolution
SX-COA01-1k = 11-2000 µS/cm0.1 µS/cm
SX-COA01-10k = 1010-20000 µS/cm1 µS/cm
  • One integrated conductivity transmitter
  • 5 m cable (or the length ordered)

4Safety, care and installation

4.1Installation

  • Submerged installation: run the cable through a rigid waterproof pipe and screw the pipe onto the upper NPT 3/4" thread of the probe, sealing the thread with PTFE tape. The top of the probe and the cable must stay dry.
  • Pipe installation: screw the probe into the pipe or tee through its NPT 3/4" thread.
  • Keep the electrode fully immersed and away from air bubbles. Do not exceed 0.6 MPa or 60 °C.
  • Wire A to A and B to B; swapped wires are the most common cause of a probe that does not answer.

4.2Care and maintenance

  • The probe needs no routine maintenance. If it fails, do not open it: contact SIGNUX.
  • An electrode that has been stored dry must be soaked in distilled water for several hours before use, to reactivate it. An electrode in frequent use may be stored in distilled water.
  • Cleaning: organic deposits come off with warm water and detergent, or with alcohol. Calcium and magnesium scale is best removed with 10 % citric acid. Clean the electrode poles only chemically or by shaking them in water: wiping them destroys the platinum black coating.
  • Calibrate before each campaign of use and, in continuous use, every three months. Adjust the frequency to the fouling and chemical deposits of the application.

5Calibration

Calibration is done over RS485, from the SIGNUX tool or from any Modbus master, with registers 0x0110 and 0x0111 written together with function code 16.

Rinse the electrode, immerse it in a standard solution of known conductivity, shake it gently to speed up the response and let it settle. When the reading is stable, write 0x0004 to register 0x0110 and the conductivity of the standard to register 0x0111, multiplied by 10 for the k = 1 probe (1-2000 µS/cm) or as it is for the k = 10 probe (10-20000 µS/cm). Example 3 in the Modbus chapter shows the exact frame.

Two related settings can also be adjusted: the temperature compensation coefficient (registers 0x0052-0x0053, 0.02 by default) for solutions with a different temperature coefficient, and the cell constant (registers 0x0054-0x0055). Both are 32-bit floating-point values, high word first. Small corrections to the reading can be made with the offsets in 0x0050 (temperature) and 0x0051 (conductivity).

6Modbus RS485 protocol

6.1Communication parameters

Encoding8-bit binary
Data bits8
ParityNone
Stop bits1
Error checkCRC (cyclic redundancy check)
Baud rate4800 bit/s by default (1200 to 115200 configurable)
Device address1 (0x01) by default

6.2Frame format

Silence before the frame≥ 4 byte times
Address code1 byte: the address of the transmitter, unique on the bus (factory default 0x01)
Function code1 byte: the operation requested by the master
Data areaN bytes: register addresses, counts and values, high byte first
Error check16-bit CRC, low byte first
Silence after the frame≥ 4 byte times

All 16-bit values are sent high byte first. The 32-bit floating-point registers (compensation coefficient and cell constant) are big-endian: high word in the lower address.

6.3Modbus registers

Register addressFunction codes (DEC)Description
0x00003 / 4Conductivity. UINT16: 10 times the value on the k = 1 probe (1-2000 µS/cm), the value itself on the k = 10 probe (10-20000 µS/cm)
0x00013 / 4Temperature (°C). INT16: 10 times the value
0x00023 / 4Salinity (ppm). UINT16
0x00033 / 4TDS (ppm). UINT16
0x00503 / 4 / 6 / 16Temperature offset. INT16, 10 times the value
0x00513 / 4 / 6 / 16Conductivity offset. UINT16, same scaling as 0x0000
0x0052 / 0x00533 / 16Temperature compensation coefficient. 32-bit float, big-endian (0.02 by default)
0x0054 / 0x00553 / 16Cell constant. 32-bit float, big-endian
0x0110 / 0x011116Calibration: write 0x0004 to 0x0110 and the standard value to 0x0111 (same scaling as 0x0000), in one frame
0x07D03 / 4 / 6 / 16Device address: 1-254 (UINT16, factory default 1)
0x07D13 / 4 / 6 / 16Baud rate: 0 = 2400, 1 = 4800, 2 = 9600, 3 = 19200, 4 = 38400, 5 = 57600, 6 = 115200, 7 = 1200 (factory default 1 = 4800)
  • UINT16: 16-bit unsigned integer
  • INT16: 16-bit signed integer

7Communication examples

Example 1: read conductivity and temperature

send frame:

Master request: 01 03 0000 0002 C4 0B

Slave address1 byte0x01
Function code1 byte0x03
Start register2 bytes0x0000
Number of registers2 bytes0x0002
CRC2 bytes0xC4 0x0B

Response frame: (For example, a device with a range of 0~2000μs/cm reads a conductivity value of 1000μs/cm and a temperature of 26.5°C)

Slave response: 01 03 04 2710 0109 30 D4

Slave address1 byte0x01
Function code1 byte0x03
Number of data bytes1 byte0x04
Conductivity2 bytes0x2710
Temperature2 bytes0x0109
CRC2 bytes0x30 0xD4
  • Conductivity calculation: 2710 (hex) = 10000 => Conductivity = 1000.0 μs/cm
  • Temperature calculation: 109H (hexadecimal)=265=>temperature=26.5℃
Example 2: set the conductivity offset

Perform numerical correction on the deviation value of the current conductivity value setting of the device whose address is 01

Sending frame: (if the current range is 0~2000μs/cm, the output conductivity value is 990, tocorrect the value to 1000, the difference is 1000-990=10, and the expansion is 10 times to100=>64H (ten Hexadecimal), the register content is written 00 64)

Master request: 01 06 0050 0064 88 30

Slave address1 byte0x01
Function code1 byte0x06
Register address2 bytes0x0050
Register content2 bytes0x0064
CRC2 bytes0x88 0x30

Response frame: (according to the MODBUS standard, the response is a mirrored message of the delivered frame)

Slave response: 01 06 0050 0064 88 30

Slave address1 byte0x01
Function code1 byte0x06
Register address2 bytes0x0050
Register content2 bytes0x0064
CRC2 bytes0x88 0x30
Example 3: calibrate with a 1413 uS/cm standard

Use the standard solution of 1413μs/ms to calibrate the device with the address 01ranging from 1 to 2000

Sending frame: write 00 04 to 0110H and 0111H respectively, and convert 1413*10 intohexadecimal to 37 32

Master request: 01 10 0110 0002 04 00043732 29 17

Slave address1 byte0x01
Function code1 byte0x10
Register address2 bytes0x0110
Register length2 bytes0x0002
Length in bytes1 bytes0x04
Register content4 bytes0x00 0x04 0x37 0x32
CRC2 bytes0x29 0x17

Response frame: (according to the MODBUS standard, the response is a mirrored message of the delivered frame)

Slave response: 01 10 0120 0002 41 fe |||

Slave address1 byte0x01
Function code1 byte0x10
Register address2 bytes0x0120
Register lengtht2 bytes0x0002
CRC2 bytes0x41 0xfe

Gallery

The SX-COA01 conductivity transmitter
Dimensions in mm: 266 overall, 212 body, 30 diameter, NPT 3/4 threads
Immersed installation
Pipe installation

Related products

Documentation

Does it fit your case?

Tell us where the SX-COA01 is going and what you need to integrate it with. We’ll tell you whether it’s the right device — and if not, which one is.

See the catalogue